ELECTROSURGICAL / CAUTERY UNIT

electrosurgical unit is a long standing work horse and provides cutting and coagulation in monopolar or bipolar outputs. The generator has audible and visual alarms, including REM™ Fault Indicator. The REM™ (Return Electrode Monitoring) is standard. 300 watts is available for output in the cut setting. A monopolar foot switch is included, a bipolar foot switch is optional as is a rolling cart. Refurbished units are available from Avante Health Solutions.

 

PRODUCT FEATURES

  • Blended Cut Modes
  • Provide flexibility through varying degrees of hemostasis. Choose one of three preset modes.
  • Simultaneous Independent Coagulation
  • Permits two surgeons to fulgurate from a single generator, for added convenience and efficacy. Accessories are activated only when keyed, reducing the potential for injury caused by inadvertent activation.
  • Low Voltage Coagulation
  • Ensures controlled, precise desiccation with less destruction of peripheral tissue, making the Force° 2 generator ideal for laparoscopic procedures.
  • REM™ Safety
  • Is guaranteed with patented adaptive REM° system, which continually monitors patient impedance levels. If a fault in the patient/return electrode contact is detected, the REM° system automatically deactivates the generator virtually eliminating the risk of burns under the return electrode. REM° safety has been proven in more than 100 million surgical procedures worldwide.

 

 

SPECIFICATIONS

Physical Weight & Dimensions

  • Height: 8 inches (20 cm)
  • Width: 13 inches (33 cm)
  • Depth: 21 inches (53 cm)
  • Weight: 23 lbs (10.4 kg)

Output Waveforms

  • Cut: 510 kHz sinusoid
  • Blend 1: 510 kHz sinusoidal bursts at 50% duty cycle recurring at 31 kHz
  • Blend 2: 510 kHz sinusoidal bursts at 37.5% duty cycle recurring at 31 kHz
  • Blend 3: 510 kHz sinusoidal bursts at 25% duty cycle recurring at 31 kHz
  • Coag: 510 kHz damped sinusoidal bursts with a repetition frequency of 31 kHz Low Voltage 510 kHz sinusoidal bursts at 25%
  • Coag: duty cycle recurring at 31 kHz
  • Bipolar: 510 kHz sinusoid Output power changes by less than 5% or 5 watts, whichever is greater, as the line voltage varies from 85-135 volts (into a 300 ohm load).
  • Low Frequency Leakage (50-60 Hz) Source current, patient leads, all outputs tied together.
  • Normal polarity, intact chassis ground < 10 µA
  • Normal polarity, ground open <100 µA
  • Reverse polarity, ground open <100 µA
  • Sink current, 140V applied, all inputs <150 µA

High Frequency Leakage

Less than 150 mA rms

Input Power Requirements

Operating range is 85 to 135 AC volts. Current is less than 8 amperes in cut and less than 4 amperes in coag.

Power Readouts

Agree with actual power into rated load to within ± 15% or 5 watts, whichever is greater.

Adaptive REM™ System

  • Measurement Frequency: 140 kHz ± 20 kHz
  • Measurement Current: 3 mA maximum
  • Acceptable Resistance Ranges:
    • REM° pad Ñ 5-135 ohms
    • Non-REM° pad Ñ less than 20 ohms
  • Acceptance range is 5-135 ohms after REM° PolyHesive° II return electrode is applied. REM° trip is initial impedance plus 40%. For example, if the initial impedance is 30 ohms, the upper level trip is approximately 42 ohms.

Cooling

Convection, no fan.

Audio Volume

The mode indicator tones are adjustable to a minimum level of 45 dB at 1 meter. The alarm tones are not adjustable and are set at 65 dB minimum at 1 meter. Also available in a 220-240 volt, 50-60 Hz configuration. Designed to meet UL and CSA specifications.